Innovative method for objective evaluations of the quality of CT systems.

In: Proceedings (World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany). Berlin: Springer, 2009. 421-424 (IFMBE Proceedings ; 25/II)
DOI Verlagsversion bestellen
An innovative method is presented that allows an objective and quantitative determination of quality parameters in computed tomography (CT). In contrast to projection radiography, the image quality parameters of a CT system is not only affected by the performance of the detector alone but additionally by the geometrical alignment of source-object-detector and the combination of numerous single raw exposure frames to one image using a specific reconstruction algorithm. Our scientific objective is to separate the effects of detector and image reconstruction algorithm on the performance of a CT. We focus here on the determination of the Modulation Transfer Function (MTF) and Noise Power Spectrum (NPS) of raw data and CT image.
Altmetric
Weitere Metriken?
Zusatzinfos bearbeiten [➜Einloggen]
Publikationstyp Artikel: Konferenzbeitrag
Schlagwörter computed tomography (CT); quality assurance; MTF; NPS; image reconstruction
ISBN 978-3-642-03878-5
Konferenztitel World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany
Konferenzband Proceedings
Quellenangaben Band: 25/II, Heft: , Seiten: 421-424 Artikelnummer: , Supplement: ,
Reihe IFMBE Proceedings
Verlag Springer
Verlagsort Berlin