Bonifacio, E.* ; Yu, L.* ; Williams, A.K.* ; Eisenbarth, G.S.* ; Bingley, P.J.* ; Marcovina, S.M.* ; Adler, K.* ; Ziegler, A.-G.* ; Mueller, P.W.* ; Schatz, D.A.* ; Krischer, J.P.* ; Steffes, M.W.* ; Akolkar, B.*

Harmonization of glutamic acid decarboxylase and islet antigen-2 autoantibody assays for national institute of diabetes and digestive and kidney diseases consortia.

J. Clin. Endocrinol. Metab. 95, 3360-7 (2010)
DOI Order publishers version
Open Access Green as soon as Postprint is submitted to ZB.
BACKGROUND/RATIONALE: Autoantibodies to islet antigen-2 (IA-2A) and glutamic acid decarboxylase (GADA) are markers for diagnosis, screening, and measuring outcomes in National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK) consortia studies. A harmonization program was established to increase comparability of results within and among these studies. METHODS: Large volumes of six working calibrators were prepared from pooled sera with GADA 4.8-493 World Health Organization (WHO) units/ml and IA-2A 2-235 WHO units/ml. Harmonized assay protocols for IA-2A and GADA using (35)S-methionine-labelled in vitro transcribed and translated antigens were developed based on methods in use in three NIDDK laboratories. Antibody thresholds were defined using sera from patients with recent onset type 1 diabetes and healthy controls. To evaluate the impact of the harmonized assay protocol on concordance of IA-2A and GADA results, two laboratories retested stored TEDDY study sera using the harmonized assays. RESULTS: The harmonized assays gave comparable but not identical results in the three laboratories. For IA-2A, using a common threshold of 5 DK units/ml, 549 of 550 control and patient samples were concordantly scored as positive or negative, specificity was greater than 99% with sensitivity 64% in all laboratories. For GADA, using thresholds equivalent to the 97th percentile of 974 control samples in each laboratory, 1051 (97.9%) of 1074 samples were concordant. On the retested TEDDY samples, discordance decreased from 4 to 1.8% for IA-2A (n = 604 samples; P = 0.02) and from 15.4 to 2.7% for GADA (n = 515 samples; P < 0.0001). CONCLUSION: Harmonization of GADA and IA-2A is feasible using large volume working calibrators and common protocols and is an effective approach to ensure consistency in autoantibody measurements.
Altmetric
Additional Metrics?
Edit extra informations Login
Publication type Article: Journal article
Document type Scientific Article
ISSN (print) / ISBN 0021-972X
e-ISSN 1945-7197
Quellenangaben Volume: 95, Issue: 7, Pages: 3360-7 Article Number: , Supplement: ,
Publisher Endocrine Society
Publishing Place Bethesda, Md.
Reviewing status Peer reviewed
Institute(s) Institute of Diabetes and Obesity (IDO)
Institute of Diabetes Research Type 1 (IDF)
Institute for Pancreatic Beta Cell Research (IPI)