PuSH - Publikationsserver des Helmholtz Zentrums München

Development and test of a new scanning geometry for Computed Tomography.

München, Technische Universität München, Fakultät für Physik, Diss., 2009, 134 S.
Verlagsversion Volltext
This work presents an alternative scanning geometry for Computed Tomography with Double Optimal Reading, called CT D’OR. With this geometry, the Radon data are acquired after the X-ray beam is collimated through a special mask, which enables an efficient data collection and may reduce the patient dose. We have demonstrated with a test-device that the scanning geometry CT D’OR acquires data for two independent and complementary reconstructions. Additionally, we have shown with a Monte Carlo simulation that the image quality attained using this scanning geometry is equivalent to the one attained by conventional geoametries.
Weitere Metriken?
Zusatzinfos bearbeiten [➜Einloggen]
Publikationstyp Sonstiges: Hochschulschrift
Typ der Hochschulschrift Dissertationsschrift
Schlagwörter Medical Imaging; Computed Tomography; scanning geometry; prototype; x-ray dose
Quellenangaben Band: De, Heft: , Seiten: 134 S. Artikelnummer: , Supplement: ,
Hochschule Technische Universität München
Hochschulort München